The Niscomed SpO2 Stimulator (MS-100) is a specialized medical device engineered for the precise testing, calibration, and validation of pulse oximeters. This essential tool ensures the accuracy and reliability of SpO2 monitoring equipment in diverse clinical environments. It is designed to simulate oxygen saturation (SpO2) levels and pulse rates, providing a reliable reference for device performance verification.


This portable stimulator features a compact and lightweight design, making it convenient for use and transport across various healthcare settings. It operates on a long-lasting, rechargeable lithium battery, enhancing its utility for extended clinical testing without constant power access.


Key Features : 

Simulates oxygen saturation (SpO2) and pulse rate for comprehensive pulse oximeter testing and calibration.
Capable of generating a wide range of simulated SpO2 levels from 35% to 100% with 1% resolution.
Simulates pulse rates ranging from 20 bpm to 250 bpm with an accuracy of 1% ± 1 bpm.
Includes built-in compatibility for 10 R-curves, such as BCI, Masimo, and Nellcor, to meet diverse testing requirements.
Features preset simulation for 24 patient states, including weak pulse, bradycardia, and hypoxia.
Supports SpO2 and pulse rate simulation under 60 Hz/50 Hz and light interference, aiding in performance evaluation under varied conditions.
Equipped with a TFT screen for adjustable brightness and a clear display of parameters.
CE marked, indicating compliance with European health, safety, and environmental protection standards.

Core Clinical FunctionSpO2 and Pulse Rate Simulation for Oximeter Calibration
Simulated SpO2 Range35%~100%
SpO2 Simulation AccuracyWhen SpO2 is 75%~100%, error is ±2% or device's accuracy (whichever is greater); when SpO2 is 74%~50%, error is ±2% or device's accuracy (whichever is greater); no definition when SpO2 is less than 50%
Simulated Pulse Rate Range20 bpm ~ 250 bpm
Pulse Rate Simulation Accuracy1% ± 1 bpm
Display TypeTFT LCD
R-Curves Supported10 (e.g., BCI, Masimo, Nellcor)
Patient States Preset24
Regulatory ApprovalCE
Power Source3.7V Rechargeable Lithium Battery
Weight0.50 KGS
twitterlinkedinfacebookinstagramyoutube