Product Description:
The RMS Medulla 201 is a state-of-the-art Two Channel BERA (Brainstem Evoked Response Audiometer) designed for precise and reliable auditory testing. It performs BERA, MLR, LLR, and P300 tests with advanced signal processing and customizable user settings. This model allows simultaneous left/right ear comparison on a single screen and supports recording up to 16 waves for accurate clinical analysis. Its intuitive interface, programmable intensity levels, and compatibility with standard PCs make it a perfect solution for audiology and neurology professionals.


Key Features

  • Supports BERA, MLR, LLR & P300 tests.
  • Two-channel configuration for simultaneous ear comparison.
  • Up to 16 waveform recordings for detailed auditory response analysis.
  • Manual/Auto Jewett Marking on waveforms.
  • Programmable intensity and latency–intensity graph plotting.
  • Hot keys for fast and easy operation.
  • Compatible with PC/Laptop systems.
  • Prints on A4-sized paper using laser or inkjet printers.
  • User-definable rate and envelope selection (Linear, Gaussian, Blackman, Hanning).
  • White noise contralateral masking from 0 dB to 80 dB nHL.

 

 

ParameterSpecification
Channels2
Sensitivity0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500 µV/div; 1, 2, 5, 10, 20 mV/div
High Cut Filter2 pole (12 dB/octave), selectable at 100, 200, 500 Hz; 1, 2, 3, 5, 10 kHz
Low Cut FilterSelectable at 0.2, 2, 20, 30, 100, 200, 500 Hz
Sweep Speeds1 to 1000 ms/div in 19 steps (1, 1.5, 2, 3, 5, etc.)
Evoked Potentials (EPs)7.5, 10, 15, 20, 30, 50, 75, 100, 150, 200, 300, 500, 750, 1000
CMRR> 100 dB
Input Impedance> 100 MΩ (common mode)
Noise< 0.5 µV rms (1 Hz to 10 kHz)
A/D Converter14-bit analog-to-digital
AveragerUp to 9999 averages per channel
Intensity Range0–110 dB nHL (30–140 dB SPL)
PresentationLeft, right, or both ears
Click Duration100 µs square wave clicks (rarefaction, condensation, alternating polarity)
Envelope OptionsLinear, Gaussian, Blackman, Hanning
MaskingWhite noise contralateral masking (0–80 dB nHL)
RateUser definable
Stimulus TypeClicks, pips
Frequency Range250 Hz to 8000 Hz
Input DeviceStimulator headphones
Computer CompatibilityWindows XP/7 (32-bit), Core2Duo or higher, 2GB RAM, 40GB HDD, 1024×768 resolution
PrintingA4 size output via laser/inkjet printer
Cart (Optional)Two-tier, heavy-duty cart with locking castors, keyboard shelf, and printer/PC tower space

 

twitterlinkedinfacebookinstagramyoutube