mb+ Disposable Skin Stapler (STP35WWH) - Pack of 10
mb+ Disposable Skin Stapler (STP35WWH) - Pack of 10: Image 1
mb+ Disposable Skin Stapler (STP35WWH) - Pack of 10: Image 2

Something off with this product?

Incorrect prices? Missing information? Delivery time too high?

4.8/5

(92)

mb+ Disposable Skin Stapler (STP35WWH) - Pack of 10

View Product Details

₹187.54/piece

(₹1,876.35)

₹10,695

82% Off

Inclusive of 5% GST

Code

Your Advantages

Brand: mb+

Country of Origin: India

What's in the Box

1 x Stapler

mb+ Disposable Skin Stapler (STP35WWH) is a preloaded, single-use surgical skin stapler with stainless steel staples and a plastic handle, used to close skin wounds in general surgery, emergency care and outpatient procedures. It addresses the problem that suturing surgical and traumatic lacerations by hand is time-consuming, prolongs operative time and can strangulate tissue when tension is uneven.

Squeezing the handle of the surgical stapler delivers individual stainless steel staples across aligned wound edges, and each staple is formed into a rectangular shape that holds the skin together until healing occurs.

In use, the stapler closes an incision with a row of staples placed at regular intervals, giving secure approximation of the skin edges.

Key Benefits

  • Ready for Immediate Wound Closure: The stainless steel staples come preloaded in the instrument, so no manual loading is needed. Closure can begin as soon as the sterile pack is opened, which keeps the procedure moving.
  • Lightweight Handle for Manual Actuation: The plastic handle serves as the handpiece, and a full squeeze of the trigger forms each staple. Its light weight suits manual operation.
  • Supplied Sterile by Ethylene Oxide: Each stapler is sterilised with ethylene oxide before it is packed. It reaches the sterile field ready for use on a single patient, with no sterilisation step needed beforehand.
  • No Cleaning or Reprocessing Needed: As a disposable skin stapler, it is built for single use. Nothing has to be cleaned or reprocessed between procedures, which removes that workload and the cross-contamination risk it carries.

The staples are made of stainless steel and the handle is made of plastic. The plastic handle forms the handpiece that the hand squeezes to deliver each staple.

Brandmb+
Model NumberSTP35WWH
Material / CompositionStainless steel (staples), plastic (handle)
Primary Use / IndicationSkin wound closure in general surgery, emergency care and outpatient procedures
SterileYes
Sterilisation MethodEthylene oxide
Single UseYes
Storage ConditionsStore in a dry, well-ventilated environment away from direct sunlight and caustic gases, with relative humidity below 80%
  1. Check the packaging and expiry date, open it in a sterile field and do not use it if the packaging is damaged.
  2. With toothed tissue forceps, bring both skin edges together and evert them slightly and evenly.
  3. Place the nose of the stapler centrally over the everted edges, perpendicular to the skin, without excessive downward pressure.
  4. Squeeze the trigger handle through a full stroke until motion stops, forming the staple into a rectangle.
  5. Release the handle fully to free the staple from the anvil before lifting the stapler.
  6. Repeat at regular intervals, typically 0.5 to 1 cm apart, until the incision is closed.
  7. Check the staple line for complete formation and haemostasis, and replace any malformed staple using a sterile staple remover.
  • Contraindications: Do not use where less than 5 mm separates the skin from underlying bone, blood vessels, nerves or viscera, or for internal tissue or visceral organ closure. Avoid with known allergy to stainless steel or its constituent metals, such as nickel or chromium, in facial, neck, hand or foot wounds needing fine suturing, and in non-linear wounds or those with severe tissue mismatch or significant gaps.
  • Possible Adverse Effects: Possible effects are incomplete staple formation or wound dehiscence, injury to underlying vessels or nerves, wound margin inversion leading to poor scarring, and surgical site infection. Strict aseptic technique lowers the infection risk.
  • Handling and Maintenance: Use on one patient only. Cleaning, reprocessing or resterilising compromises mechanical integrity and sterility.
  • Storage: Keep in a dry, well-ventilated place away from direct sunlight and caustic gases, with relative humidity below 80%.
  • Disposal: Discard used staplers as bio-medical sharps or plastic waste under the Bio-Medical Waste Management Rules.