Dora Hollow Fiber Dialyzer B-14P is a high-flux polyethersulfone (PES) capillary dialyzer designed for effective removal of uremic toxins during dialysis treatment. uses the same PES fibre, 200 µm inner diameter with a 40 µm wall, potted in polyurethane with a 1.4m²; surface area to support efficient solute removal and ultrafiltration. Its biocompatible construction, low priming volume, and sterile presentation support patient safety and efficient clinical workflows.


Engineered for precision, the dialyzer features biocompatible construction that minimizes the risk of allergic reactions and ensures patient safety optimizing the removal of both small and middle molecular solutes during dialysis sessions. P series has a polycarbonate housing sterilised by radiation, with non-pyrogenic blood and dialysate pathways.


Hollow Fiber Dialyzer with smaller surface areas and priming volumes are prone to suit lighter adults and patients prone to intradialytic hypotension. However, more clearance is well-suited for under-dialysed or heavy patients. Dialyzer's clearance rate indicates clinical effectiveness of a hemodialysis treatment. Essentially, these clearance targets determine how efficiently a dialyzer removes metabolic waste products (like urea) and excess fluids from a patient's blood during a scheduled session.


Dora Dialyzer is a single-use dialyzer which lasts efficiently for a four-hour session. They are designed for single-use only, reuse significantly drops its filtering capacity, risks protein buildup, and increases infection or biocompatibility hazards. Since high-flux alternatives enable longer & quick fluid extraction, they are widely recommended for longterm care cases. Hollow Fiber Dialyzers are known for offering a large surface area for efficient filtration than others.



Features:

  • Advanced Hollow Fiber Technology: Supports efficient solute and uremic toxin removal.
  • Polyethersulfone Membrane: Provides biocompatibility and stable ultrafiltration characteristics.
  • Low Priming Volume: Helps minimize the blood volume required to initiate dialysis.
  • Biocompatible Construction: Minimizes interaction with blood components.
  • Fibres: polyethersulfone hollow fibres, 200 µm inner diameter, 40 µm wall, 1.3 m² effective area.
  • Potting: polyurethane.
  • Housing and end caps: polypropylene on the PF series; polycarbonate on the P series.
  • Sterilisation: radiation; blood and dialysate pathways sterile and non-pyrogenic.
ModelB-14P
Membrane TypeHollow Fiber
Flux TypeHigh-Flux
TypeHollow fibre capillary hemodialyzer
Fiber Diameter / Wall200 µm / 40 µm
Membrane MaterialPolyethersulfone (PES)
Effective Surface Area1.4 m²
SterilizedYes
IndicationAcute / Chronic Renal Failure
Maximum TMP500 mmHg (66 kPa)
PottingPolyurethane (PU)
ConnectorsLuer-lock blood port & couplers
Priming Volume85ml
Urea cl. QB 300251
  1. Check: Ensure to verify these points before opening, check if pouch is intact, within expiry, model matches the prescription. 
  2. Connect: Clamp the dialyzer vertically with the arterial (red) end down for priming then connect the arterial bloodline to the lower blood port and the venous line to the upper port by luer-lock; connect the dialysate lines so dialysate flows counter-current to blood.
  3. Prime and rinse: Use heparinised normal saline to prime, then tap the housing to clear air from the fiber bundle and fill the dialysate compartment, then check for leaks. Invert the dialyzer so the arterial end is up once priming is complete.
  4. Observe: Follow applicable clinical protocols and manufacturer instructions throughout treatment. Monitor arterial and venous pressures, TMP and the patient throughout. Stop and replace the dialyzer on a blood-leak alarm, visible clotting or a hypersensitivity reaction.
  • Do not use a dialyzer that has been frozen, wetted or dropped.
  • Do not refrigerate or freeze before use.
  • Do not use a dialyzer with compromised package, broken housing, or expired shelf life.
  • Compatibility with the Dora Hemodialysis Model series must be confirmed before use.
  • Aseptic techniques should be followed during setup to avoid contamination.
  • Patient vitals should be monitored closely throughout the dialysis process.