*Images Are For Representation Purpose Only Actual Product May Differ

Defibtech Lifeline Auto Fully Automatic Defibrillator DCF-2310-EN (AED)

Availability: In stock

Quick Overview

Features:
• The Lifeline AUTO is a fully automatic defibrillator designed to analyze heart rhythm and automatically deliver a shock—without intervention—if needed.

• The Lifeline AUTO includes all mission critical features necessary to provide the most advanced treatment for Sudden Cardiac Arrest. Yet it’s so simple and unintimidating to use that even non-medical personnel can effectively save lives.

• The Lifeline AUTO was developed by experienced multidisciplinary engineering teams and incorporates state-of-the-art digital signal processing techniques and advanced ECG analysis algorithms. The device exceeds the American Heart Association performance recommendations—giving the user confidence the correct therapy is being delivered.

• The Lifeline AUTO defibrillator uses advanced biphasic technology—including the most studied biphasic shock waveform—and automatically adjusts the shock delivery to the person’s individual needs.

• For first response professionals like police, fire and EMS, the Lifeline AUTO is standard equipment. For schools, offices, malls, factories, and public places, it’s becoming as vital as a fire extinguisher
Technical specifications:
• Type: Fully Automatic external defibrillator
• Model : DDU-120A, DDU-120E
• Video Prompts: Full motion video On- screen video help
• Waveform: Biphasic Truncated Exponential ( Impedance compensated )
• Energy
Adult: 150 Joules
Child/ Infant: 50 Joules
(Nominal into 50 Ohm load )
• Voice Prompts: Extensive voice prompts guide user through operation of the unit and CPR
• CPR Pacing: Metronome
• Controls: Light On/Off button
• Indicators:
- Check pads
- Do not touch patient
• Charge Time : DBP-2800: Less than 6 seconds DBP-1400: Less than 9 Seconds
GMDN: 48046
Check Delivery Availability

View Price
Defibrillation Monitoring
• Model:
- Adult: DDP-100
Child/ Infant - DDP-200P
- Surface Area:
• Surface Area:
103 cm² ( nominal, each pad )
50cm² (nominal, each pad )
• Pad Placement:
• Adult : Anterior/ Anterior
• Child/Infant: Anterior/ Posterior
• Type: Pre-Connected, single-use, non-polarized, disposable , self-adhesive, electrodes with cable connector.
• Cable Length: ( typical) 48 in (122 cm )


Event Documentation:
• Internal Event Record: Critical ECG segments and rescue event parameters are recorded( greater than 60 minutes ) and can be downloaded to a removable data card.
• PC-Based Event Review : ECG with event display, and audio playback with when available.
• Removable Storage : ( Operational ) Up to 30 hours of ECG and event data storage ( no audio option ) or up to 3 hours of audio ( audio option ). ECG and event storage on a removable data card. Actual length of storage is dependent card capacity.

Patient Analysis System:
• Patient Analysis: Automatically evaluates patient impedance for proper pad contact. Monitors signal quality and analyzes patient ECG for shockable / non-shockable rhythms.
• Sensitivity/ Specificity: Meets AAMI-DF-80 specifications and AHA recommendations.

Environmental:
• Temperature:
Operating: 0 to 50°C(32 to 122°F)
Standby : 0 to 50°C (32 to 122°F)
• Relative Humidity:
Operating/ Standby : 5%- 95% ( non-Condensing)
• Altitude:
-500 to 15,000 ft (-150 to 4500 m) per MIL-STD-810F 500. 4Procedure II
• Vibration:
Ground (MIL-STD-810F 514.5 Category20)
Helicopter (RTCA/DO-160D, Section 8.8.2, Cat R.Zone2, Curve G )
Jet Aircraft (RTC/DO-160D Section , Cat H , Zone 2 , Curves B & R )
• Shock/Drop Abuse Tolerance
MIL-STD-810F 516.5 Procedure IV 48 inches (1 meters, any edge. Corner, or surface , in standby mode.)
• Sealing/ Water Resistance: IEC 60529 class IP54 ; Dust protected ( battery pack installed )
• ESD : IEC 61000-4-2: 2009(Open air up to 15kV or direct contact up to 8kV )
• EMC ( Emission ): EN 60601-1-2:2001+ A1:2006, limits
EN 55011:1998 + A1 : 1999 + A2: 2002
Group 1 Level B, method
• EMC ( Immunity ): EN 60601-2-4:2003, limits
EN 61000-4-3:2002 Level
3(10V/m),method


Patient Analysis System:
• Patient Analysis: Automatically evaluates patient impedance for proper pad contact. Monitors signal quality and analyzes patient ECG for shockable/ non-shockable rhythms.
• Sensitivity/ Specificity: Meets AAMI-DF-80 specifications and AHA recommendations

Battery Pack:
• Model DBP- 2800
- Power: 15V, 2800 mAh
- Capacity: 300 shocks or 16 hours continuous operation
- Standby Life: 7 years

• Model DBP- 1400
- Power: 15V, 1400 mAh
- Capacity: 125 shocks or 8 hours continuous operation
- Power: 15V, 1400 mAh
- Standby Life: 5 years
• Type: Lithium/ Manganese Dioxide Disposable, Rechargeable
• Low Battery Indicators: Visible
Audible

Self Tests:
• Automatic : Automatic daily, weekly, monthly and quarterly circuitry tests
• Battery Insertion : System integrity test on battery insertion
• Pad Presence: Pads preconnected tested daily
• User-Initiated: Unit and battery pack system test initiated by three user
• Status Indication: Visual and audible indication of unit status

Physical:
• Size:
8.5 x 11.8 x 2.7 Inches
(22 x 30 x 7 cm )
• Weight( Approximate )
With DBP-1400: 4.2 lbs(1.9kg)
With DBP-2800: 4.4 lbs (2.0 kg)